Chapter 4: Test and Calibration Procedures
ABL77 Service Manual
Calibration Lines Procedure,
Continued
Calibration lines
procedure
(continued)
Step Action
5.
If a sensor cassette is…
Attached
The analyzer will remove any fluid, then prompt
the operator to remove the sensor cassette. The
analyzer will sense when the cassette has been
removed and will automatically proceed to the
next screen.
not attached
The analyzer will automatically proceed to the
next step.
RESULT:
The following screen will appear.
6.
Ensure there is a cal pack on the analyzer to collect the fluids exiting
the waste line.
7.
Introduce 1 cc of Cleaning Solution (REF: 943-906) into the waste
drain (black opening located directly below the inlet probe when a
cassette is in place).
Continued on next page
4-6
Rev.
F