Chapter 8: Maintenance
ABL77 Service Manual
Long-Term Storage,
Continued
Long-term
storage
(continued)
Step Action
15.
Remove the syringe and pull back on the plunger to fill the syringe
with air. Replace the syringe on the cassette luer.
16.
•
Press
Cal 1
.
•
While
Cal1 Open
is displayed on the screen, gently introduce air
into the luer. This will remove residual fluids from the Cal 1 line.
•
Press
Continue
.
17.
Remove the syringe and fill it with deionized water. Replace the
syringe on the cassette luer.
18.
•
Press
Cal 2
.
•
While
Cal2 Open
is displayed on the screen, gently introduce the
deionized water into the luer. It will exit the Cal 2 manifold port.
•
Press
Continue
.
19.
Remove the syringe and pull back on the plunger to fill the syringe
with air. Replace the syringe on the cassette luer.
20.
•
Press
Cal 2
.
•
While
Cal2 Open
is displayed on the screen, gently introduce air
into the luer. This will remove residual fluids from the Cal 2 line.
•
Press
Continue
.
21.
•
Press
Shutdown
(followed by
Yes
) to turn off the analyzer.
•
Wipe any external analyzer surfaces with a damp cloth.
8-12
Rev.
F