ABL77 Service Manual
Chapter 1: Introduction
Test Equipment and Tools
Required
equipment and
tools
To carry out the procedures for servicing the analyzer, the following test
equipment and tools are required: Parts marked with “N/A” in the REF column are
to be purchased locally.
REF Description
Use
902-6XX SC
i
sensor cassette
Testing analyzer after repair
944-069 Calibration solutions pack (cal pack)
Testing analyzer after repair
902-578 Electronic zeroing fixture
Zeroing sensor channels
913-741 Manager setup disk
Custom setup of system features
920-721
5
/
64
” ball point Allen wrench
Various applications
920-722
1
/
16
” ball point Allen wrench
Lower module
920-723
7
/
64
” ball point Allen wrench
Lower module
920-726
5
/
16
” open-end wrench
Electronics module
920-727
5
/
16
” nut driver
Electronics and Lower modules
920-728 ¼” nut driver
Electronics module
920-729 Valve board calibration cassette
Calibrating the Hct and SC circuit
991-260 Loctite 401 Adhesive
Securing the handle to the housing
991-264 Silicone adhesive, RTV medical grade
Securing cables to prevent disconnect
991-265 Loctite 222 Threadlocker
Securing assorted screws and nuts
905-674 Sample path obstruction tool kit
Various applications
943-906 Cleaning solution
Cleaning the waste drain
N/A
Deionized water (or sterile water)
Various applications
N/A
Soft cloth, gauze or other absorbent material
Various applications
N/A
Protective attire (gloves, safety glasses, and lab
coat)
Various applications
N/A
Flat head screwdriver
Lower module
N/A
Small flat head screwdriver
Adjusting the calibration settings
N/A
Phillips head screwdriver
Various applications
N/A
Needle nose pliers
Lower module
N/A
Scissors or equivalent trimming tool
Various applications
N/A
AT style keyboard
Entering system commands
N/A Digital
multimeter
Testing
electronics
N/A
Digital voltmeter (or multimeter)
5 volt power output setting
N/A
3½” formatted diskettes
Downloading data files
N/A
Index card or equivalent
Block the opto sensor
Rev. F
1-5