Chapter 7: Re-Assembly
ABL77 Service Manual
Analog Board
Introduction
Follow these steps to install the analog board. The board must be prepared for
proper operation with the ABL77 Analyzer.
NOTE:
The valve board and analog board are a matched set for the Hct and SC
circuits and must be calibrated when either is replaced. This also applies when
replacing the lower module or electronics module, which contains these boards.
Calibration is not required when these boards are only removed and put back into
the same analyzer. Refer to Chapter 4 for instructions on performing this
procedure.
Continued on next page
7-58