ABL77 Service Manual
Chapter 6: Replacements
Manifold,
Continued
Installing the
new manifold
Follow these instructions to install a new manifold.
Step Action
1.
Connect the Cal1 port tubing from the new manifold to the adapter
connected to the N.C. valve port of
L1
. Make certain the tubing is
completely installed.
2.
Connect the Cal2 port tubing from the new manifold to the adapter
connected to the N.C. valve port of
L2
. Make certain the tubing is
completely installed.
3.
Connect the new manifold waste tubing to the bulkhead connector
(item
1
in figure 6-17) on the printer chassis.
4.
Make certain that all of the cables and tubing are not pinched.
41449 / B
41477
SERVICE BOARD ABL70-77
ASSY REV :
K1
J5
R2
R18
REVISION
J1
0
PN 902
-81
1
J1
J9
SENSOR BOSS
SENSOR
LUER
J6
J4
J8
J7
J3
12
1
13
24
J2
2587
J6
J5
J4
J9
J8
J7
L2
L1
L3
J10
C
a
l2
W
ast
e
C
al
1
S H
LTP3345B-C576
A B C
PULL
PUSH
MA
DE
IN
MAL
A
Y
S
IA
ST
EP
PER
M
O
T
O
R
5:
VD
C
OHM
S
/COI
L:
6.
2
5
ST
EP
A
N
GL
E:
7
.5
L8270
1-
P1
L3
COM N.
C.
N.
O.
L1
COM N.
C.
N.
O.
L2
COM N.
C.
N.
O.
1
Figure 6-17
Continued on next page