ABL77 Service Manual
Chapter 3: Troubleshooting
Reference Table,
Continued
Error #
Message
Interpretation
Corrective Action
1600
Temperature Errors
1601
Sample: Temperature Not
Stable
Temperature stabilization
criteria not met during
sample analysis
See
Chapter 4, Test and
Calibration Procedures
to
investigate the heater circuit
1602
Flush: Temperature Not
Stable
Temperature stabilization
criteria not met during
sample flush
See
Chapter 4, Test and
Calibration Procedures
to
investigate the heater circuit
1603 QC
Sample:
Temperature
Not Stable
Temperature stabilization
criteria not met during QC
sample analysis
See
Chapter 4, Test and
Calibration Procedures
to
investigate the heater circuit
1604
QC Flush: Temperature Not
Stable
Temperature stabilization
criteria not met during QC
sample flush.
See
Chapter 4, Test and
Calibration Procedures
to
investigate the heater circuit
1605
Cal 1: Temperature Not
Stable
Temperature stabilization
criteria not met during Cal 1
measurement
See
Chapter 4, Test and
Calibration Procedures
to
investigate the heater circuit
1606
Cal 2: Temperature Not
Stable
Temperature stabilization
criteria not met during Cal 2
measurement
See
Chapter 4, Test and
Calibration Procedures
to
investigate the heater circuit
1700
Upload Errors
1701
Connection timed out
Maximum time exceeded for
uploading data to remote
host computer
Check connections. Verify
the remote host system is
working properly.
1702
Receiver not ready
Remote host system has not
acknowledged it is ready to
receive data.
Wait a few moments then
retry. If problem persists,
verify the remote host
system is working properly
1703
Receiver cancelled
Remote host system has
cancelled the
communication.
Wait a few moments then
retry. If problem persists,
check connections. Verify
the remote host system is
working properly.
Continued on next page
3-23