Chapter 4: Test and Calibration Procedures
ABL77 Service Manual
Calibration Lines Procedure,
Continued
Calibration lines
procedure
(continued)
Step Action
17.
Press
Continue
.
RESULT:
The following screen will appear.
18.
•
Refill the syringe with deionized or sterile water and flush the Cal2
line in a similar manner, pressing
Cal2
to open a pathway between
the cassette luer and the Cal2 manifold luer.
•
Repeat as necessary.
•
Press
OK
when flushing is complete.
19.
•
Re-attach the cal pack and sensor cassette (if one is currently
installed).
•
Press
OK
.
RESULT:
The system will return to the
Options
menu.
20.
Perform a manual calibration to ensure proper function prior to sample
analysis.
4-10
Rev.
F