Chapter 6: Replacements
ABL77 Service Manual
Valve Board Tubing,
Continued
Replacing the
tubing
(continued)
Step Action
NOTE:
Reference figures 6-13, 6-14 and 6-15 for assembly details
and references.
7.
Using the “A” tubing, with a tubing plug attached, connect the open
tubing end to the N.O. port of valve L3.
8.
a) Using the “B” tubing, connect the length of clear tubing to the
N.C. port of valve L3 with the Tee connector and silicone tubing
facing the waste drain
b) Connect one end of the silicone tubing to the waste drain
c) Connect the other end of the silicone tubing to the printer chassis
port
9.
a) Connect one end of the “C” tubing to the COM port of valve L3
b) Connect the other end to the N.O. port of valve L1
10.
Using one “D” tubing, with a tubing adapter attached, connect the
open tubing end to the N.C. port of valve L1. The adapter end of this
tube will be connected later to the Cal2 port tubing from the manifold
during analyzer re-assembly.
11.
a) Connect one end of the “E” tubing to the COM port of valve L1
b) Connect the other end to the N.O. port of valve L2
12.
Using the other “D” tubing, with a tubing adapter attached, connect
the open tubing end to the N.C. port of valve L2. The adapter end of
this tube will be connected later to the Cal1 port tubing from the
manifold during analyzer re-assembly.
13.
a) Using the “F” tubing, connect the (shorter) thick tubing end to the
COM port of valve L2.
b) Connect the thin tubing end to the sensor cassette luer
Continued on next page
6-24
Rev.
F