Chapter 4: Test and Calibration Procedures
ABL77 Service Manual
Waste Line,
Continued
Waste line
procedure
(continued)
Step Action
9.
Press
OK
.
10.
If the analyzer has a sensor cassette:
currently installed…
not currently installed…
The analyzer will prompt to replace
the sensor cassette. Then press
OK
.
The analyzer will perform a
mandatory 2 pt cal then return to the
Main Menu
.
The analyzer will
immediately exit to the
Main Menu
.
4-28
Rev.
F