Chapter 3: Troubleshooting
ABL77 Service Manual
Reference Table,
Continued
Error #
Message
Interpretation
Corrective Action
1311
Calibration required before
analysis.
Maximum allowable time
between calibrations has
been reached.
Perform manual calibration
or allow auto calibration to
complete.
1312
QC Level 1 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 1 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 1
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 2 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 2 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 2
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 3 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 3 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 3
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
QC Level 4 out of range
At least one parameter fell
outside the acceptable range
for the most recent
BG/Lytes Level 4 QC
measurement
•
Perform a manual
calibration and repeat the
BG/Lytes Level 4
measurement
•
If results remain out of
range inactivate this
parameter or replace the
sensor cassette.
Continued on next page