Chapter 8: Maintenance
ABL77 Service Manual
Valve Maintenance,
Continued
Valve
maintenance
(continued)
Step Action
16.
Remove any excess water from the syringe. Pull back on the syringe
plunger to fill the syringe with air.
17.
Attach the syringe to the cassette luer and press
Cal1
to repeat this
process using air.
18.
Press
Continue
.
19.
Refill the syringe with deionized water and flush the Cal2 line in a
similar manner, pressing
Cal2
to open a pathway between the cassette
luer and the Cal2 manifold luer.
20.
Refill the syringe with deionized water and press
Cal2
to repeat this
process one more time.
21.
Remove the syringe from the analyzer.
22.
Remove any excess water from the syringe. Pull back on the syringe
plunger to fill the syringe with air.
23.
Attach the syringe to the cassette luer and press
Cal2
to repeat this
process using air.
24.
Press
OK
when flushing is complete.
25.
Re-attach the cal pack and sensor cassette.
NOTE:
DO NOT re-install the disposables through the software if
they had been previously installed.
26.
•
Press
OK
.
•
The system will return to the
Options
menu.
Continued on next page
8-4
Rev.
F