Chapter 8: Maintenance
ABL77 Service Manual
Long-Term Storage,
Continued
Long-term
storage
(continued)
Step Action
3.
If a sensor cassette is… then…
attached
the analyzer will remove any fluid then
prompt you to remove the sensor
cassette. The analyzer will sense when
the cassette has been removed and will
automatically proceed to the next
screen.
not attached
the analyzer will automatically proceed
to the next screen.
4.
There are three components of the ABL77 that become contaminated
with sample residue with use. These components are the sensor
cassette, the waste line and the waste pouch within the cal pack. The
sensor cassette and cal pack can be discarded as biohazardous waste.
5.
To decontaminate the waste line ensure there is a cal pack in place to
collect waste fluids. Introduce 1 cc of Cleaning Solution (REF: 943-
906) into the waste drain (the black opening located directly below the
inlet probe when a cassette is in place).
6.
•
Activate the waste pump by pressing
Pump
to carry this fluid
through the waste line and into the cal pack.
•
Press
Continue
.
7.
•
Introduce 1 cc of deionized water into the waste drain.
•
Press
Pump
to clear the fluid.
•
Press
Continue
.
8.
•
Remove the cal pack
•
Press
Continue
.
Continued on next page
8-10
Rev.
F