Model: MTS200 (1U)
Doc. Ref. no. : - m08/om/201
Issue no.: 03
Page 183 of 195
User’s Manual
If
mas
TER
T-Sync
Model MTS200 is Power ON in Unlock conditions, the unit will provide time output
depending on the data of its internal RTC clock time which is available with battery backup (refer section
8). However, if the provided battery backup to RTC is discharged due to very long Power OFF period of
this device, at Power ON conditions in unlock conditions; all outputs of device will have factory set time
value and not the correct time. Once unit gets locked, all outputs will get proper time data.
mas
TER
T-Sync
Model MTS200 device enter Holdover mode, when unit goes into Unlock condition from
Lock condition and thereafter provides time output depending on the internal clock crystal accuracy. The
accuracy of all time outputs (including 1PPS output) of unit will degrade depending on the duration during
which unit is in Holdover mode and also on the internal clock crystal frequency accuracy. If the Unit again
enter the Lock condition from Unlock condition, all the time outputs will become accurate as per UTC
time. Holdover mode conditions do not exist if unit gets power reboot while unit was in Unlock condition.
mas
TER
T-Sync
Model MTS200 outputs will regain its accuracy only when unit gets in lock condition once
after Power ON.
15 Options
mas
TER
T-Sync
Model MTS200 model can be configured for Optional Power Supply.