Model: MTS200 (1U)
Doc. Ref. no. : - m08/om/201
Issue no.: 03
Page 67 of 195
User’s Manual
Here fdc2:7142:0:1eba:8cff:fee5:c115 is the Global IPv6 address of the MTS200. It can be
viewed on LCD, Webserver, SNMP and on console based configuration utility.
user
: root
password
: MTS200LAMBDA
Note: Operator can also use “putty” software for windows based system for establishing telnet or
ssh communication with MTS200 unit. Please refer section 10.1
for steps to use “putty” software
for establishing telnet or ssh session.
After operator have entered the login of MTS200, then give below command to start the application
program:
Command
: /usr/sbin/start
And then press enter, start menu will be display as shown in below figure
Figure 9-3 Console based Program main menu layout
Above menu is the start menu of application code, which list out the entering code for different applicable
modes such as “GEN” for General settings, “ETH” for device network settings etc. After entering any one
of given mode and pressing “enter” will make device enter into that particular mode.
In all modes described in main menu image as per figure, user can check the current configurations using
“View settings” option in respective modes and also user can change the current settings by using
“configure settings” option in their respective modes. All parameters command number are obtained from
left most column when “View settings” option is used.
Whenever operator do configuration,
operator can save the set parameters using “0” command and then
press “y” or “Y” (yes) when prompted for saving. Whenever user is in middle of configuration mode option
and want to check the command options of other parameters of same mode, user can press
“H” for help.
This list out all command values of other parameters of same mode.