Model: MTS200 (1U)
Doc. Ref. no. : - m08/om/201
Issue no.: 03
Page 147 of 195
User’s Manual
mas
TER
T-Sync
Model MTS200 device provides 1PPS output at every 1 second through its BNC terminal
on rear panel of unit. This is TTL signal of 0(low level) to 5V (high level) value. The Pulse width of 1PPS
signal is 20% duty cycle i.e. 200 milliseconds (high level) and 800 milliseconds (low level).
12.2.2 Event Output (PPM/PPH)
mas
TER
T-Sync
Model MTS200 device is equipped with the standard feature of providing event output at
every PPM (Pulse per Minute) / PPH (Pulse per Hour) with fix pulse width of 1 second. This event
provides pulse output according to configured time interval i.e. Minute / Hour through front panel keypad
or console based configuration utility or webserver or snmp. Please refer section 9 for method of
configure standard event output through console based configuration utility and to configure through
keypad on device or section 13.3 for snmp or section 13.4 for webserver. Refer technical specification
section 4 for electrical characteristics of event output.
12.2.3 Additional Event Outputs (Programmable Pulse Outputs)
mas
TER
T-Sync
Model MTS200 device is equipped with the (optional) feature of providing 1 to 4
additional event outputs. These events provide pulse output according to configured time interval and ON
time. Each event time can be configured with time interval ranging from 1 sec to 86400 seconds and
pulse ON time (pulse width) from min. 50 milliseconds to max. 50% of configure time interval of that
particular event in terms of milliseconds through front panel keypad or console based configuration utility
or webserver or snmp. Please refer section 9 for method of configure additional event outputs through
console based configuration utility and keypad or section 13.3 for snmp or section 13.4 for webserver on
device. Refer technical specification section 4 for electrical characteristics of additional event outputs.