![Keysight Technologies N5171B EXG User Manual Download Page 93](http://html1.mh-extra.com/html/keysight-technologies/n5171b-exg/n5171b-exg_user-manual_1962194093.webp)
Keysight EXG and MXG X-Series Signal Generators User’s Guide
77
Optimize Performance
Using the USB Pass Through Commands
Using the USB Pass Through Commands
The USB pass through SCPI commands are used remotely and or to program your system setup
and power meter sensor setup.
This section applies to the following USB power sensors:
— U2040 X-Series
— U2020 X-Series
— U8480 Series
— U2000 Series
—
This section contains the following:
—
Table 4-1, “SCPI Pass Through Commands”
—
Procedure
If the power sensor is configured to return peak power, then the MXG
displays peak power (i.e., Use the
:fetch?
and
:measure?
power sensor
SCPI commands, which returns a value based on the configuration). Refer
to the X-Series Signal Generators SCPI Command Reference.
Table 4-1
SCPI Pass Through Commands
:SYSTem:PMETer[1]|2:PASSthrough <"scpiCommand">
:SYSTem:PMETer[1]|2:PASSthrough? <"scpiQuery">
:SYSTem:PMETer[1]|2:PASSthrough:ENABle 0|1
:SYSTem:PMETer[1]|2:PASSthrough:ENABle?
:SYSTem:PMETer[1]|2:PASSthrough:TIMeout <valueInSeconds>
:SYSTem:PMETer[1]|2:PASSthrough:TIMeout?
Step
Substeps
Results/Notes
1.
Setup equipment.
2.
Enable the pass
through capability in
the MXG.
a.
Enter:
:SYSTem:PMETer:PAS
Sthrough:ENABle 1
3.
Query the instrument
and verify pass
through status.
a.
Enter:
:SYSTem:PMETer:PAS
Sthrough:ENABle?
The instrument should return a
1
, indicating that the
pass through feature has been enabled.