238
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Digital Operation (Option 653/655/656/657)
Scaling a Waveform
Setting Waveform Runtime Scaling
Runtime scaling scales the waveform data during playback; it does not affect the stored data. You
can apply runtime scaling to either a segment or sequence, and set the scaling value either while
the ARB is on or off. This type of scaling is well suited for eliminating DAC over–range errors.
Runtime scaling adjustments are not cumulative; the scaling value is applied to the original
amplitude of the waveform file. There are two ways to save the runtime scaling setting: by using the
save function (
) and by saving the setting to the file header (
). Saving to the file
header saves the value with the waveform file, saving with the Save function stores the value as the
current instrument setting.
Use this example to learn how to scale the currently selected waveform.
1.
Select the waveform to which you want to apply scaling:
a.
Press
Mode
>
Dual ARB
>
Select Waveform
.
b.
Highlight the desired waveform (segment or sequence).
c.
Press
Select Waveform
.
2.
Play the selected waveform: Press
ARB Off On
until On highlights.
3.
Set the Waveform Runtime Scaling value:
a.
Press
ARB Setup
>
Waveform Runtime Scaling
.
b.
Enter a scaling value.
The signal generator automatically applies the new scaling value to the waveform. There is
no single value that is optimal for all waveforms. To achieve the maximum dynamic range,
use the largest scaling value that does not result in a DAC over–range error.
c.
Press
Return
.