
90
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Optimize Performance
Using Flatness Correction
The
UF
annunciator appears in the
AMPLITUDE
area of the display, and the correction data in the
array is applied to the RF output.
Recalling and Applying a User Flatness Correction Array
The following example assumes that a user flatness correction array has been created and stored. If
Example: A 500 MHz to 1 GHz Flatness Correction Array with 10 Correction
1.
Preset the signal generator.
2.
Recall the desired User Flatness Correction file:
a.
Press
AMPTD
>
More
>
User Flatness
>
Configure Cal Array
>
More
>
Preset List
>
Confirm
Preset
.
b.
Press
Load/Store
.
c.
Highlight the desired file.
d.
Populate the user flatness correction array with the data contained in the selected file:
Press
Load From Selected File
>
Confirm Load From File
.
The user flatness correction array title displays
User Flatness:
Name of File
.
3.
Apply the correction data in the array to the RF output
:
Press
Return
>
Flatness Off On
to
On
.