164
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Analog Modulation (Option UNT)
Using an External Modulation Source
Using Wideband AM
Wideband AM uses the I input of the I/Q modulation system. When the wideband AM is turned on,
the I/Q is turned on and the I/Q source is set to external. If the I/Q is turned off or the I/Q source is
set to internal, then the wideband AM turns off.
Figure 6-2
Wideband AM Softkey Menu
Setting the Wideband AM
1.
Set up and enable the desired modulation type.
2.
Press
AM > AM Path 1 2 WB
to
WB.
For frequencies between 9 kHz and 5 MHz, Wideband AM turns off.
AM > AM Path 1 2 WB
Enables and disables the
wideband AM feature.
Note: If the I/Q is turned off or
the I/Q source is set to
internal, then the wideband
AM turns off.
For details on each key, use key help
as described on
When the Wideband AM is enabled, these fields are active.