Keysight EXG and MXG X-Series Signal Generators User’s Guide
163
Analog Modulation (Option UNT)
Using an External Modulation Source
Using an External Modulation Source
Removing an External Source DC Offset
To eliminate an offset in an externally applied FM or
Φ
M signal, perform an external DC calibration
(Ext DC Cal).
1.
Set up and turn on the desired modulation.
2.
Press
FM/
Φ
M
>
FM Source > Ext1 or Ext2 > More
>
Setup FM Source.
3.
Ensure that Ext Coupling is set to DC.
- If not, press
Ext Coupling DC AC
until DC is highlighted.
4.
Press
Ext DC Cal
.
This begins the calibration.
Performing the calibration with a DC signal applied removes any deviation caused by the DC signal,
and the applied DC level becomes the new zero reference point. When you disconnect the DC
signal, perform the calibration again to reset the carrier to the correct zero reference.
Rear panel inputs are described on
AM, FM or
Φ
M inputs