226
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Digital Operation (Option 653/655/656/657)
Clipping a Waveform
Clipping a Waveform
Digitally modulated signals with high power peaks can cause intermodulation distortion, resulting
in spectral regrowth that can interfere with signals in adjacent frequency bands. The clipping
function enables you to reduce high power peaks by clipping the I and Q data to a selected
percentage of its highest peak, thereby reducing spectral regrowth.
—
—
How Peaks Cause Spectral Regrowth
—
How Clipping Reduces Peak–to–Average Power
—
—
Configuring Rectangular Clipping
Figure 9-8
Clipping Softkeys
Mode
>
Dual ARB
>
More
Available only when
clipping type = |I|, |Q|
For details on each key, use key help
as described on
.