324
Keysight EXG and MXG X-Series Signal Generators User’s Guide
BERT (Option UN7)
Bit Error Rate Tester–Option UN7
Bit Error Rate Tester–Option UN7
The bit error rate test (BERT) capability allows you to perform bit error rate (BER) analysis on digital
communications equipment. This enables functional and parametric testing of receivers and
components including sensitivity and selectivity.
Block Diagram
When measuring BER, a clock signal that corresponds to the unit under test (UUT) output data
must be input to the BER CLK IN connector. If the clock is not available from the UUT, use the DATA
CLK OUT signal from the X-Series baseband modulator. Refer to
for information
about these connections.
Figure 12-1
Clock Gate Function
When you use the clock gate function, the clock signal to the BER CLK IN (rear panel BB TRIG 1)
connector is valid only when the clock gate signal to the BER GATE IN connector is ON.
Press the
Clock Gate Off On
softkey to toggle the clock gate function off and on.The
Clock Gate
Polarity Neg Pos
softkey sets the input polarity of the clock gate signal supplied to the rear panel
BER GATE IN connector. When you select
Pos
(positive), the clock signal is valid when the clock
gate signal is high; when you select
Neg
(negative), the clock signal is valid when the clock gate
signal is low.
The following figure shows an example of the clock gate signal.