338
Keysight EXG and MXG X-Series Signal Generators User’s Guide
BERT (Option UN7)
Verifying BERT Operation
Measurement Example Using Custom Digital Modulation (Requires Option
431)
The following steps set up the signal generator for a BERT measurement using Custom Digital
Modulation.
1.
Refer to
and make the following connections on the signal generator’s rear panel.
—
DATA OUT
(
Aux I/O
connector pin 33) to BER DATA IN (BNC connector labeled EVENT 1).
—
DATA CLK OUT
(
Aux I/O
connector pin 7) to BER CLK IN (BNC connector labeled BB
TRIG1).
2.
Press the
Preset
hardkey. This configures the signal generator to a pre-defined state.
3.
Press the
Mode
hardkey.
4.
Press
Real-Time Custom > Modulation Setup >
Modulation Type > QPSK.
5.
Press
Return > Return > Return > Data
>
PN Sequence
>
More > PN9
.
6.
Press
Return > More > More > Data/Clk/Sync Signal Polarity Setup > Data/Clk/Sync Rear
Outputs On
.
7.
Press the
Aux Fctn
hardkey.
8.
Press
Digital Bus Connectivity PXB/BERT N5102A to N5102A > Confirm Changes > BERT >
BERT Off On
to On >
BER Display > BER Display % Exp
to % >
Display Update Cycle End Cont
to Cont.
The following steps configure BERT measurement parameters.
9.
Press
Return > Configure BERT
>
Total Bits
to 100000 >
Enter > More > Special Pattern Ignore
Off On
to Off.
10.
Press
BERT Resync Off On
to Off >
Return > Pass/Fail Off On
to Off.
11.
Press
Return
>
I/O Setup
>
Gate Control > Gate Off On
to Off.
12.
Press
Return > Polarity Setup
>
Clock Polarity Pos Neg
to Neg >
Data Polarity Pos Neg
to Pos.
13.
Press
Return
>
Return
>
Configure Trigger > Cycle Count 0 > Enter
.
14.
Press
Bit Delay Off On
to Off >
BERT Trigger
to Trigger Key.
15.
Press the
Trigger
shows the signal generator’s front-panel display after
completion of the these steps.