Keysight EXG and MXG X-Series Signal Generators User’s Guide
317
Digital Signal Interface Module (Option 003/004)
Operating the N5102A Module in Input Mode
Figure 10-22
Data Setup Softkey Menu with Parallel Port Configuration
2.
Press the
Data Type
softkey.
In this menu, select the data type to be either filtered (
Samples
) or unfiltered (
Pre-FIR
Samples
). The selection is dependent on the test needs and the device under test. However if
the device being tested already incorporates FIR filters, the
Pre-FIR Samples
selection should
be used to avoid double filtering. Refer to
, for more information.
3.
Select the data type that is appropriate for the test needs.
4.
Press the
Numeric Format
softkey.
Inactive for a serial port
Frame polarity is active
configuration
for a serial port configuration
Only available when the N5102A
digital module is
turned on and using input mode
Only available when
Data Type is
Pre-FIR Samples