Keysight EXG and MXG X-Series Signal Generators User’s Guide
75
Optimize Performance
Using the Dual Power Meter Display
13.
On the signal generator: Press Channel B to On and then back to Off again. This initializes the
signal generator to the external power meter.
14.
Press
Return
>
Zero Sensor
A diagnostic dialog box is displayed each time an external power meter is being used and the
Zero Sensor or Calibrate Sensor softkey is pressed (refer to
).
Verify the power sensor is connected to the 50 MHz reference of the power meter.
Figure 4-10
Diagnostic Dialog Box for Channel B
15.
Press
Done
The
Running Calibration(s)
bar is displayed:
Zeroing Sensor Please wait...
.
16.
After Running Calibration(s) bar disappears: Press
Calibrate Sensor
Diagnostic Dialog box is displayed that prompts for verifying the connection of the power
sensor to the power meter 50 MHz reference (refer to
Figure 4-11
Diagnostic Dialog Box for Calibration
For details on
each key, use key
help as described
on
.
For details on
each key, use key
help as described
on
.