320 Keysight EXG and MXG X-Series Signal Generators User’s Guide
Baseband Operating Mode—Primary, BERT, or N5102A
To set the Baseband Operating Mode to BERT
To set the Baseband Operating Mode to BERT
In this mode, all installed and licensed signal generator features as well as all
BERT features are fully supported, but N5102A features are not available.
Figure 11-2
Changing to the BERT Baseband Operating Mode
1.
Press
Aux Fctn
.
2.
Press
Baseband Operating Mode
.
(Default selection is
Primary
.)
3.
Press
BERT
. (Press the
Return
key to cancel the change.)
4.
After reading the above caution and taking any needed actions,
press
Confirm Change to BERT Operating Mode
or
Return
to cancel.
This confirmation selects the BERT ARB FPGA image.
All BERT features are fully supported, but N5102A is not available in this
mode. The BERT ARB FPGA image contains all functionality and features
from the B.01.65 PXB Input FPGA image. For backwards compatibility,
selecting the BERT Baseband Operating Mode sets the I/Q Digital Bus
Connectivity to PXB.
The BERT annunciator is displayed indicating that the
BERT Baseband Operating Mode is currently selected.
(For more information on annunciators, see
.)
Changing the Baseband Operating Mode selection will cause all files in
volatile w
aveform memory to be lost. Backup any files you want to keep before
confirming the change.
See
“Working with Files” on page 56
.