430
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Troubleshooting
Internal Media Data Storage
Amplitude Does Not Change in List or Step Sweep
Verify that sweep type is set to amplitude (Amptd); the amplitude does not change when the sweep
type is set to frequency (Freq) or waveform.
Internal Media Data Storage
Instrument State Saved but the Register is Empty or Contains the Wrong
State
If the register number you intended to use is empty or contains the wrong instrument state, recall
register 99. If you selected a register number greater than 99, the signal generator automatically
saves the instrument state in register 99.
See also
“Working with Instrument State Files” on page 62
USB Media Data Storage
Instrument Recognizes USB Media Connection, but Does Not Display Files
If the USB media works on other instruments or computers, it may simply be incompatible with the
signal generator; try a different USB media. Refer to
http://www.keysight.com/find/mxg
for
details on compatible USB media.
Preset
The Signal Generator Does Not Respond
If the signal generator does not respond to a preset, the instrument may be in remote mode, which
locks the keypad.
To exit remote mode and unlock the preset keys, press
Local Cancel/(Esc).
Pressing Preset Performs a User Preset
This behavior results from the use of a backward–compatible SCPI command. To return the signal
generator to normal use, send the command
:SYST:PRESet:TYPE NORM
.
For information on SCPI commands, refer to the
SCPI Command Reference
.