
402
Keysight EXG and MXG X-Series Signal Generators User’s Guide
Custom Digital Modulation (Option 431)
Modifying a FIR Filter Using the FIR Table Editor
5.
Press
Filter Symbols
>
8
>
Enter
.
6.
Press
Generate
.
7.
Press
Display Filter > Display Impulse Response
(refer to
Figure 16-39
Impulse Response Display
8.
Press
Return
.
Modifying the Coefficients
1.
Using the front panel arrow keys, highlight coefficient 15.
2.
Press
0
>
Enter
.
3.
Press
Display Impulse Response
.
Figure 16-40
Impulse Response Display with Modified Coefficients
Refer to
. The graphic display can provide a useful troubleshooting
tool (in this case, it indicates that a coefficient value is missing, resulting in an improper
Gaussian response).
4.
Press
Return
.
The actual oversample ratio during modulation is automatically selected
by the instrument. A value between 4 and 16 is chosen dependent on the
symbol rate, the number of bits per symbol of the modulation type, and
the number of symbols.
For details on each key, use key
help as described on
For details on each key, use key
help as described on
.