Keysight EXG and MXG X-Series Signal Generators User’s Guide
Basic Digital Operation (Option 653/655/656/657)
Using Waveform Markers
Using the RF Blanking Marker Function
While you can set a marker function (described as
Marker Routing
on the softkey label in the
Marker Utilities menu) either before or after setting the marker points (
), setting a marker
function before you set marker points may change the RF output. RF Blanking includes ALC hold
(described on
, note
regarding unleveled power). The signal generator blanks the
RF output when the marker signal goes low. This example is a continuation of the previous example,
1.
Using the factory–supplied segment
SINE_TEST_WFM
, set Marker 1 across points 1
−
180
(
).
2.
From the
Marker Routing
softkey menu, assign RF Blanking to Marker 1:
In the second Arb menu (
), press
Marker Utilities
>
Marker Routing
>
Pulse/RF
Blank
>
Marker 1
.
Marker Polarity = Positive
When marker polarity is positive (the default
setting), the RF output is blanked during the
off
marker points.
200
180
Point 1
Marker
Segment
RF Signal
200
180
Point 1
Marker
Segment
Marker Polarity = Negative
When marker polarity is negative, the
RF output is blanked during the
on
marker
points
≈
3.3V
0V
≈
3.3V
0V
RF Signal
RF
Signal
RF