Keysight EXG and MXG X-Series Signal Generators User’s Guide
Analog Modulation (Option UNT)
Configuring the LF Output (Option 303)
DC
selects a DC voltage level as the LF output BNC source.
Configuring the LF Output with an Internal Modulation Source
In this example, the internal FM is the LF output source. See
.
Configuring the Internal Modulation as the LF Output Source
1.
Press
Preset
.
2.
Press the
FM/
Φ
M
hardkey.
3.
Press
FM Dev
>
75
>
kHz
.
4.
Press
More > Setup FM Source > FM Rate
>
10
>
kHz
.
5.
Press
Return > Return > FM Off On
.
You have set up the FM signal with a rate of 10 kHz and 75 kHz of deviation. The
FM
annunciator is
activated indicating that you have enabled frequency modulation.
Configuring the Low Frequency Output
1.
Press the
LF Out
hardkey.
2.
Press
LF Out Amplitude
>
3
>
V
.
3.
Press
LF Out Off On
.
You have configured the LF output signal for a 3 volt sine wave (default wave form) output which is
frequency modulated using the
Int Monitor
source selection (default source).
The
LF Out Off On
softkey controls the operating state of the LF output. However when the LF
output source selection is
Int Monitor
, you have three ways of controlling the output. You can
use the modulation source (AM, FM, or
Φ
M) on/off key, the LF output on/off key, or the
Mod
On/Off
softkey.
The
RF On/Off
hardkey does not apply to the LF OUTPUT connector.
Internal modulation (
Int Monitor
) is the default LF output source.