Appendix D: Model 2450 in a Model 2400 application
Model 2450 Interactive SourceMeter® Instrument Reference Manual
D-8
2450-901-01 Rev. B/September 2013
Model 2400 command
:CALCulate2:LIMit[1]:COMPliance:FAIL
:CALCulate2:LIMit[1]:COMPliance:FAIL?
Model 2450 command
Not available
Notes
Model 2400 command
:CALCulate2:LIMit[1]:COMPliance:SOURce2
:CALCulate2:LIMit[1]:COMPliance:SOURce2?
Model 2450 command
Not available
Notes
Model 2400 command
:CALCulate2:LIMit[1]:FAIL?
Model 2450 command
Not available
Notes
Model 2400 command
:CALCulate2:LIMit[1]:STATe
:CALCulate2:LIMit[1]:STATe?
Model 2450 command
Not available
Notes
To disable source limits, set the limit value to the maximum allowed by the
instrument
Model 2400 command
:CALCulate2:LIMit<x>:FAIL?
Model 2450 command
:CALCulate2:<function>:LIMit<y>:FAIL?
(on page 6-15), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
Notes
Note that
:CALCulate2:LIMit4:FAIL?
is not supported (related to contact
check, which is not provided on Model 2450).
Note that this only available for two limits in the Model 2450.
Model 2400 command
:CALCulate2:LIMit<x>:LOWer[:DATA]
:CALCulate2:LIMit<x>:LOWer[:DATA]?
Model 2450 command
:CALCulate2:<function>:LIMit<y>:LOWer[:DATA]
(on page 6-16), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
Notes
Note that this only available for two limits in the Model 2450.
Model 2400 command
:CALCulate2:LIMit<x>:LOWer:SOURce2
:CALCulate2:LIMit<x>:LOWer:SOURce2?
Model 2450 command
Not available
Notes
Model 2400 command
:CALCulate2:LIMit<x>:PASS:SOURce2
:CALCulate2:LIMit<x>:PASS:SOURce2?
Model 2450 command
Not available
Notes
Model 2400 command
:CALCulate2:LIMit<x>:STATe
:CALCulate2:LIMit<x>:STATe?
Model 2450 command
:CALCulate2:<function>:LIMit<y>:STATe
(on page 6-17), where
<function>
is
VOLTage[:DC]
,
CURRent[:DC]
, or
RESistance
and
<y>
is
1
or
2
.
Notes
Note that this only available for two limits in the Model 2450.
To disable source limits, set the limit value to the maximum allowed by the
instrument.