Section 3: Functions and features
Model 2450 Interactive SourceMeter® Instrument Reference Manual
3-94
2450-901-01 Rev. B/September 2013
Figure 104: SynchronousA input trigger
Output characteristics:
•
In addition to trigger events from other trigger objects, the TSP commands
trigger.digout[
N
].assert()
and
trigger.tsplinkout[
N
].assert()
release the line
if the line is latched low. The pulse width is not used.
•
The physical line state does not change until all drives (internal and external) release the line.
•
Action overruns occur if the internal drive is not latched low and a source event is received.