
Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-87
Details
This command determines how the measurement range is selected.
When this command is set to off, you must set the range. If you do not set the range, the instrument remains at
the range that was selected automatically.
When this command is set to on, the instrument automatically goes to the most sensitive range to perform the
measurement. The instrument sets the range when a measurement is requested.
If a range is manually selected through the front panel or a remote command, this command is automatically set
to off.
Example
smu.measure.func =
smu.FUNC_DC_CURRENT
smu.measure.autorange = smu.ON
Set the measurement function to current.
Set the range to be set automatically.
Also see
(on page 2-101)
(on page 8-117)
smu.measure.autorangehigh
This attribute reads the highest measurement range that is used when the instrument selects the measurement
range automatically.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW) Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
Resistance: 2e5
Usage
highRange
= smu.measure.autorangehigh
smu.measure.autorangehigh =
highRange
highRange
The highest voltage or resistance measurement range that is used when the range
is set automatically:
•
Voltage: 2e-2 to 200 volts
•
Resistance: 20 to 2e8 ohms
Details
This command is only available for voltage and resistance measurements. It can be written to and read for
resistance measurements. For voltage measurements, it can only be read.
For voltage measurements, the upper limit is controlled by the voltage limit.
Resistance measurements only: You can use this command when automatic range selection is enabled to put an
upper bound on the range that is used for resistance measurements.
The upper limit must be more than the lower limit.
If the lower limit is equal to the upper limit, automatic range setting is effectively disabled.