Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-65
:SOURce[1]:CONFiguration:LIST:STORe
This command stores the active source settings into the named configuration list.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Saved settings
Not applicable
Usage
:SOURce[1]:CONFiguration:LIST:STORe <name>, <point>
<name>
A string that represents the name of a source configuration list
<point>
A specific configuration point in the configuration list
Details
Use this command to store the active settings to a configuration point in a configuration list. If you do not include
the <point> parameter, the configuration point will append to the end of the list.
Refer to
Instrument settings stored in a source configuration list
(on page 3-38) for information about the settings
this command stores.
Example
:SOURce:CONF:LIST:CRE "biasLevel"
:SOURce:FUNC VOLT
:SOURce:VOLT:LEV 5
:SOURce:CONF:LIST:STORE "biasLevel"
Create a configuration list named
biasLevel
.
Set the source function to voltage and the source
voltage level to 5 volts.
Store the configuration list and append it to the
end of the
biasLevel
configuration list.
Also see
:SOURce[1]:CONFiguration:LIST:CREate
(on page 6-61)