Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-124
2450-901-01 Rev. B/September 2013
:TRACe:TRIGger
This command makes readings and stores them in a buffer.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRACe:TRIGger
:TRACe:TRIGger <bufferName>
<bufferName>
A string that indicates the reading buffer; the default buffers
(
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no
buffer is specified, this parameter defaults to
defbuffer1
Details
Makes readings and stores them in the specified buffer.
Example
TRACe:MAKE "MyBuffer", 100
TRACe:TRIG "MyBuffer"
TRACe:TRIG "MyBuffer"
TRACe:TRIG "MyBuffer"
TRACe:TRIG "MyBuffer"
TRACe:TRIG "MyBuffer"
TRACe:DATA? 1,5, "MyBuffer", rel
Create a buffer called
MyBuffer
with a maximum size of
100.
Take readings and store them in
MyBuffer
.
Recall the relative time when the data points were
measured for the first five readings in the buffer.
Example output:
0.000000,0.408402,0.816757,1.208823,1.6175
29
Also see
(on page 6-114)
TRIGger subsystem
The commands in this subsystem configure and control the trigger operations, including the trigger
model.
:INITiate[:IMMediate]
This command starts the trigger model.
Type
Affected by
Where saved
Default value
Command
Not applicable
Not applicable
Not applicable
Usage
:INITiate[:IMMediate]
Also see
(on page 3-95)