Section 3: Functions and features
Model 2450 Interactive SourceMeter® Instrument Reference Manual
3-42
2450-901-01 Rev. B/September 2013
If you want to recall a point on a source configuration list, under
Source
, select
Config List
.
3. Choose
Select List
. A menu of available configuration lists is displayed.
4. Select
MyMeasList
. The configuration points in the list display.
5. Select the second configuration point.
6. Select
List to System
.
7. Select
HOME
to return to the operating display.
Viewing configuration list contents
For example, use the following procedure to view configuration point 2 from
MyMeasList
.
Using the front panel to view configuration list contents:
1. Press the
MENU
key.
2. Under
Measure
, select
Config List
. The MEASURE CONFIGURATION LISTS screen is
displayed.
If you want to view a source configuration list, under
Source
, select
Config List
.
3. Choose
Select List
. A menu of available configuration lists is displayed.
4. Select
MyMeasList
. The configuration points are displayed.
5. Select the second configuration point.
6. Select
View Details
.
7. When you are finished, select
OK
.
8. Select
HOME
to return to the home screen.
Deleting a configuration point
For example, use the following procedure to delete configuration point 4 from
MyMeasList
.
Using the front panel to delete a configuration point:
1. Press the
MENU
key.
Under
Measure
, select
Config List
. The MEASURE CONFIGURATION LISTS screen is
displayed.
If you want to delete a point from a source configuration list, under
Source
, select
Config List
.
2. Choose
Select List
. A menu of available configuration lists is displayed.
3. Select
MyMeasList
. The configuration points are displayed.
4. Select the second configuration point.
5. Select
Delete From
.
6. Select
HOME
to return to the operating display.