Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-47
display.changescreen()
This function changes which front-panel screen is displayed.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
display.changescreen(s
creenName
)
screenName
The screen to display:
•
Home screen:
display.SCREEN_HOME
•
Source swipe screen:
display.SCREEN_SOURCE_SWIPE
•
Data trend swipe screen:
display.SCREEN_PLOT_SWIPE
•
User display swipe screen:
display.SCREEN_USER_SWIPE
•
Buffer statistics swipe screen:
display.SCREEN_STATS_SWIPE
•
Settings swipe screen:
display.SCREEN_SETTINGS_SWIPE
•
Graph screen:
display.SCREEN_GRAPH
•
Data sheet screen:
display.SCREEN_DATASHEET
Example
display.clear()
display.changescreen(display.SCREEN_USER_SWIPE)
display.settext(display.TEXT1, "Batch A122")
display.settext(display.TEXT2, "Test running")
Clear the User Display screen and switch
to display the User Display screen.
Set the first line to read "Batch A122"
and the second line to display "Test
running".
Also see
(on page 8-52)