Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-131
Details
This command can only return data for one configuration point. To get data for additional configuration points,
resend the command and specify different configuration points.
Refer to
Instrument settings stored in a source configuration list
(on page 3-38) for a complete list of source
settings that the instrument stores in a source configuration list.
Example
print(smu.source.configlist.query("MyScrList", 2))
Returns the TSP commands that represent the settings in configuration point 2.
Also see
(on page 3-33)
smu.source.configlist.create()
Instrument settings stored in a source configuration list
(on page 3-38)
smu.source.configlist.recall()
This command recalls a specific configuration point in a source configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
Not applicable
Not applicable
Usage
smu.source.configlist.recall(
name
,
point
)
name
A string that represents the name of a source configuration list
point
A specific configuration point in the configuration list
Details
Use this command to recall the settings stored in a specific configuration point on a specific configuration list. If
you do not specify a point when you send the command, it recalls the settings stored in the first configuration
point on the specified configuration list.
Example
smu.source.configlist.recall("MySourceList")
Since a point was not specified, this
command recalls configuration point 1
from a configuration list named
MySourceList
.
smu.source.configlist.recall("MySourceList", 5)
Recalls configuration point 5 in a
configuration list named
MySourceList
.
Also see
(on page 3-33)
smu.source.configlist.create()