:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
This command clears the results of the limit test.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:CALCulate2:<function>:LIMit<Y>:CLEar[:IMMediate]
<function>
The function to which this setting applies:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<y>
The limit that the setting applies to, 1 or 2
Details
Use this command to clear the test results of limit
Y
when the limit auto clear command is disabled. Both the high
and low test results are cleared.
To avoid the need to manually clear the test results for a limit, enable the auto clear command.
If you send this command without the
<function>
parameter, it will set the limit for all functions.
Example
calc2:curr:lim1:clear
Clear the results for limit test 1.
Also see
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
(on page 6-16)
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]
(on page 6-18)