Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-69
:SOURce[1]:<function>:HIGH:CAPacitance
This command enables or disables high capacitance feature.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Source configuration list
Save settings
Source configuration list
0 (OFF)
Usage
:SOURce[1]:<function>:HIGH:CAPacitance <b>
:SOURce[1]:<function>:HIGH:CAPacitance?
<function>
The source function to which this setting applies:
•
Current:
CURRent
•
Voltage:
VOLTage
<b>
Turn high capacitance feature on:
ON
or
1
Turn high capacitance feature off:
OFF
or
0
Details
When the instrument is measuring low current and is driving a capacitive load, you may see overshoot, ringing,
and instability. You can enable the high capacitance feature to minimize these problems.
High-capacitance settings apply when operating using the 10 nA through the 100 mA current ranges. When
operating using the 1 A range, the high-capacitance setting will not affect the instrument rise time or current
measure settling time.
Example
SOUR:CURR:HIGH:CAP ON
Turn the high capacitance feature on when
sourcing current.
Also see
(on page 4-21)