trigger.model.setblock() — trigger.BLOCK_CONFIG_RECALL
This attribute recalls the system settings that are stored in a measure or source configuration list.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_CONFIG_RECALL,
configurationList
)
trigger.model.setblock(
blockNumber
, trigger.BLOCK_CONFIG_RECALL,
configurationList
,
index
)
blockNumber
The sequence of the block in the trigger model
configurationList
The configuration list to recall
index
The point in the configuration list to recall; default is 1
Details
When the trigger model reaches a configuration recall building block, the settings in the specified configuration
list are recalled.
You can restore a specific set of configuration settings in the configuration list by defining the index.
Example
trigger.model.setblock(3, trigger.BLOCK_CONFIG_RECALL, "measTrigList", 5)
Configure trigger block 3 to load index point 5 from the configuration list named
measTrigList
.
Also see