Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-149
Trigger events
Event description
Event constant
No trigger event
NONE
Trigger event blender
<n>
(1 to 2), which combines
trigger events
BLENder<n>
Trigger timer
<n>
(1 to 4) expired
TIMer<n>
Source limit condition occurs
SLIMit
Example
:TRIGger:BLOCk:WAIT 9, DISP
Set trigger model block 9 to wait for a user to
press the TRIGGER key on the front panel
before continuing.
Also see
(on page 3-96)
:TRIGger:DIGital<n>:IN:CLEar
This command clears the trigger event on a digital input line.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:DIGital<n>:IN:CLEar
<n>
Digital I/O trigger line (1 to 6)
Details
The event detector of a trigger enters the detected state when an event is detected. For the specified trigger line,
this command clears the event detector, discards the history, and clears the overrun status
(
sets the overrun
status to
false
).
For this command to function as expected, make sure you configure the trigger type and line state of the digital
line for use with the trigger model (use the digital line mode command).
Example
:TRIG:DIG2:IN:CLE
Clears the trigger event detector on I/O line 2.
Also see
(on page 6-19)
(on page 3-62)
:TRIGger:DIGital<n>:IN:OVERrun?
(on page 6-151)