Model 2450 Interactive SourceMeter® Instrument Reference Manual
Appendix D: Model 2450 in a Model 2400 application
2450-901-01 Rev. B/September 2013
D-11
FETCh?
Model 2400 command
:FETCh?
Model 2450 command
(on page 6-3)
Notes
Can choose different buffers and which buffer elements to access from the
buffer for Model 2450.
FORMat subsystem
Model 2400 command
:FORMat:BORDer
:FORMat:BORDer?
Model 2450 command
(on page 6-29)
:FORMat:BORDer?
Notes
Model 2400 command
:FORMat[:DATA]
:FORMat[:DATA]?
Model 2450 command
(on page 6-30)
:FORMat[:DATA]?
Notes
Model 2400 command
:FORMat:ELEMents[:SENSe[1]]
:FORMat:ELEMents[:SENSe[1]]?
Model 2450 command
Not available
Notes
In the Model 2450, format elements are specified with the
READ?
,
FETCh?
,
MEASure?
, and
TRACe:DATA?
commands with each use instead of using a
global setting for all commands. The elements may be unique for each
command and are unique each time the command is processed.
Model 2400 command
:FORMat:ELEMents:CALCulate
:FORMat:ELEMents:CALCulate?
Model 2450 command
Not available
Notes
Model 2400 command
:FORMat:SOURce2
:FORMat:SOURce2?
Model 2450 command
Not available
Notes
Model 2400 command
:FORMat:SREGister
:FORMat:SREGister?
Model 2450 command
Not available
Notes
OUTPut subsystem
Model 2400 command
:OUTPut[1]:ENABle[:STATe]
:OUTPut[1]:ENABle[:STATe]?
Model 2450 command
Not available
Notes