Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-44
2450-901-01 Rev. B/September 2013
[:SENSe[1]]:<function>:AVERage[:STATe]
This command enables or disables the averaging filter for measurements of the selected function.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
0 (OFF)
Usage
[:SENSe[1]]:<function>:AVERage[:STATe] <state>
[:SENSe[1]]:<function>:AVERage[:STATe]?
<function>
The measurement function to which this setting applies:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<state>
The filter status; set to one of the following values:
•
Disable the averaging filter:
0
or
OFF
•
Enable the averaging filter:
1
or
ON
Details
This command enables or disables the averaging filter. When this is enabled, the measurements for the selected
measurement function are averaged as set by the filter count and filter type.
If you send this command without the
<function>
parameter, it sets the state of the averaging filter
for all functions.
Example 1
CURR:AVER:COUNT 10
CURR:AVER:TCON MOV
CURR:AVER ON
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 2
RES:AVER:COUNT 10
RES:AVER:TCON MOV
RES:AVER ON
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Example 3
VOLT:AVER:COUNT 10
VOLT:AVER:TCON MOV
VOLT:AVER ON
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Also see
(on page 4-22)
[:SENSe[1]]:<function>:AVERage:COUNt
(on page 6-43)
[:SENSe[1]]:<function>:AVERage:TCONtrol
(on page 6-45)