Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-101
Details
The filter count is the number of readings that are acquired and stored in the filter stack for the averaging
calculation. The larger the filter count, the more filtering that is performed.
This command is set for the selected function.
Example
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.filter.count = 10
smu.measure.filter.type = smu.FILTER_MOVING_AVG
smu.measure.filter.enable = smu.ON
Set the measurement function to
current.
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Also see
(on page 4-22)
(on page 8-101)
(on page 8-102)
smu.measure.filter.enable
This attribute enables or disables the averaging filter for the selected measurement function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.OFF
Usage
filterState
= smu.measure.filter.enable
smu.measure.filter.enable =
filterState
filterState
The filter status:
•
Disable the filter:
smu.OFF
or
0
•
Enable the filter:
smu.ON
or
1
Details
This command enables or disables the averaging filter. When this is enabled, the measurements for the selected
measurement function are averaged as set by the filter count and filter type.
Example
smu.measure.func = smu.FUNC_DC_CURRENT
smu.measure.filter.count = 10
smu.measure.filter.type = smu.FILTER_MOVING_AVG
smu.measure.filter.enable = smu.ON
Set the measurement function to
current.
Set the averaging filter type to moving
average, with a filter count of 10.
Enable the averaging filter.
Also see
(on page 4-22)
(on page 8-100)