Section 8: TSP command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
8-100
2450-901-01 Rev. B/September 2013
smu.measure.displaydigits
This attribute determines the number of digits that are displayed for measurements on the front panel for the
selected function.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
smu.DIGITS_5_5
Usage
digits
= smu.measure.displaydigits
smu.measure.displaydigits =
digits
digits
6½ display digits:
smu.DIGITS_6_5
5½ display digits:
smu.DIGITS_5_5
4½ display digits:
smu.DIGITS_4_5
3½ display digits:
smu.DIGITS_3_5
Details
This command affects how the reading for a measurement is displayed on the front panel of the instrument. It
does not affect the number of digits returned in a remote command reading. It also does not affect the accuracy
or speed of measurements.
The display digits setting is saved with the function setting, so if you use another function, then return to the
function for which you set display digits, the display digits setting you set previously is retained.
To change the number of digits returned in a remote command reading, use
format.asciiprecision
.
Example
smu.measure.func = smu.FUNC_DC_VOLTAGE
smu.measure.displaydigits = smu.DIGITS_6_5
Set the measurement function to voltage
with a front-panel display resolution of 6.5.
Also see
(on page 8-62)
smu.measure.filter.count
This attribute sets the number of measurements that are averaged when filtering is enabled.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Function change
Configuration script
Measure configuration list
10
Usage
filterCount
= smu.measure.filter.count
smu.measure.filter.count =
filterCount
filterCount
The number of readings required for each filtered measurement (1 to 100)