trigger.model.load() — Config List
This function loads a predefined trigger model configuration that uses source and measure configuration lists.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
trigger.model.load("ConfigList",
measureConfigList
,
sourceConfigList
)
trigger.model.load("ConfigList",
measureConfigList
,
sourceConfigList, delay
)
trigger.model.load("ConfigList",
measureConfigList
,
sourceConfigList
,
delay
,
readingBuffer
)
measureConfigList
A string that contains the name of the measurement configuration list to
use
sourceConfigList
A string the contains the name of the source configuration list to use
delay
The delay that will occur before the measurement is made; default is 0
readingBuffer
The name of the reading buffer, which may be a default buffer
(
defbuffer1
or
defbuffer2
) or a user-defined buffer; default is
defbuffer1
Details
This trigger model template incorporates a source configuration list and measure configuration list.
You must set up the configuration lists before loading the trigger model.
You can also set a delay and reading buffer.
Also see
None