Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 6: SCPI command reference
2450-901-01 Rev. B/September 2013
6-161
:TRIGger:LOAD:CONFiguration:LIST
This command loads a predefined trigger model configuration that uses source and measure configuration lists.
Type
Affected by
Where saved
Default value
Command only
Not applicable
Not applicable
Not applicable
Usage
:TRIGger:LOAD:CONFiguration:LIST <measureConfigList>, <sourceConfigList>
:TRIGger:LOAD:CONFiguration:LIST <measureConfigList>, <sourceConfigList>, <delay>
:TRIGger:LOAD:CONFiguration:LIST <measureConfigList>, <sourceConfigList>, <delay>,
<readingBuffer>
<measureConfigList>
The configuration list that contains the measurement settings to be used
for this trigger model
<sourceConfigList>
The configuration list that contains the source settings to be used for this
trigger model
<delay>
The delay time before the measurement (seconds); default is 0
<readingBuffer>
A string that indicates the reading buffer; the default buffers (
defbuffer1
or
defbuffer2
) or the name of a user-defined buffer; if no buffer is
specified, this parameter defaults to
defbuffer1
Details
This trigger model template incorporates a source configuration list and measure configuration list.
You must set up the configuration lists before loading the trigger model.
You can also set a delay and reading buffer.
Example
TRIG:LOAD:CONF:LIST "MyMeasList", "src config"
Load the configuration list trigger model, using the measurement configuration list
MyMeasList
and the
source configuration list
src config
.
Also see
None