Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-18
2450-901-01 Rev. B/September 2013
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]
This command specifies the upper limit for a limit test.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
1.00
Usage
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA] <n>
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]?
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? DEFault
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? MINimum
:CALCulate2:<function>:LIMit<Y>:UPPer[:DATA]? MAXimum
<function>
The function to which this setting applies:
•
Current:
CURRent[:DC]
•
Resistance:
RESistance
•
Voltage:
VOLTage[:DC]
<Y>
The limit test that the upper limit applies to, 1 or 2
<n>
The value of the upper limit (
−
9.11 to +9.11)
Details
This command sets the high limits for the limit tests for the selected measurement function. When limit testing is
enabled for this limit, the instrument generates a fail indication when the measurement value is more than this
value.
If you send this command without the
<function>
parameter, it will set the limit for all functions.
Example
CALC2:VOLT:LIM1:LOW 0.25
CALC2:VOLT:LIM1:UPP 2.5
CALC2:VOLT:LIMIT1:STAT ON
CALC2:VOLT:LIMIT1:FAIL?
Set lower limit 1 for voltage to 0.25 volts.
Set upper limit 1 for voltage to 2.5 volts.
Enable limit 1 testing for voltage.
Check the test results.
Also see
:CALCulate2:<function>:LIMit<Y>:LOWer[:DATA]
(on page 6-16)