Section 2: General operation
Model 2450 Interactive SourceMeter® Instrument Reference Manual
2-116
2450-901-01 Rev. B/September 2013
Set the power-on configuration using a SCPI command:
Send the command:
:SYSTem:POSetup <name>
Where
<name>
is:
•
RST
: Power-up to *RST defaults
•
SAV0
: Power-up to setup stored at memory location 0
•
SAV1
: Power-up to setup stored at memory location 1
•
SAV2
: Power-up to setup stored at memory location 2
•
SAV3
: Power-up to setup stored at memory location 3
•
SAV4
: Power-up to setup stored at memory location 4
Using a TSP command:
Save the script that you want to use as the power-on default to be
autoexec
. For example, to save
the commands that are presently in the instrument to be the power-on defaults, send the command:
createconfigscript("autoexec")
Reset the instrument
You can reset many of the commands to their default values. For detail on what gets reset, see
(on page 4-24). Default values are also listed in the individual command descriptions.
If you are connected to a TSP-Link system, resetting the instrument resets all TSP-Link enabled
instruments on the TSP-Link system.
Using the front panel:
1. Press the
MENU
key.
2. Under System, select
Manage
.
3. Select
System Reset
.
4. The commands are reset and a confirmation message is displayed.
Using TSP commands:
Send the command:
reset()
If the instrument is connected to a TSP-Link system and you are using TSP commands, you can
reset only the local instrument by sending
localnode.reset()
instead of
reset()
.
Using SCPI commands:
Send the command:
*RST